Heating Element: High-quality heating wire, lightweight fiber material with low thermal capacity, excellent insulation properties, fast heating, energy-saving, time-saving, and easy environmental temperature control.
Control Method: Imported single-loop intelligent temperature controller with PID control, auto-tuning of parameters.
Number of Temperature Control Zones: 1
Alarm Protection: Over-temperature, circulating fan failure, etc., with audio-visual alarm protection and automatic power cutoff. After a high-temperature alarm cuts off the power, the heating power can only be manually restored when the temperature drops to the normal operating range before restarting.
Temperature Control Accuracy: ±1°C
Temperature Uniformity: ≤±2°C
Fan Power: 1/4 HP motor, high-temperature resistant, low noise
Insulation Protection: Independently processed extended shaft core for the fan
Impeller Structure: Multi-blade design
Airflow Channel Design: Unique design with porous stainless steel walls, ensuring no dead spots in circulation.
Facility Requirements
1 .Environmental conditions: Temperature 0 - 40ºC, humidity ≤ 80% RH, no corrosive gas, no strong airflow disturbance. 2.Ventilation system: Non-contact connection to the user's exhaust system, with an exhaust capacity greater than 20 m³/h. 3 .Floor: Level, without significant vibration, load-bearing capacity > 500 Kg 4 .Power conditions: Capacity greater than 6 KVA, three-phase five-wire, voltage 220/380V, frequency 50Hz. Live wire: yellow, green, red; neutral wire: blue; earth wire: yellow-green. This power line is a dedicated line, please distribute power separately. Power sources for precision instruments and large power conversion equipment such as welding machines, electric furnaces, metal cutting machine tools, etc., should not be connected to this distribution system. 5. Installation site: 1800×1200mm×1500mm (L×W×H), installation area greater than 2 m²