1. Application:
The HRF180-05 Type Rapid Annealing Furnace is primarily designed for processes such as drying, curing, or debinding of semiconductors, electronic components, or materials.
2. Technical Specification:
2.1. Rated Temperature: 400°C;
2.2. Maximum Temperature: 450°C;
2.3. Furnace Chamber Size: 600 × 500 × 600mm (W × D × H);
2.4. Loading Device: Drawer-style tray;
2.5. Furnace Chamber Material: SUS304;
2.6. Heating Element: Custom-made squirrel-cage heater;
2.7. Temperature Stability: ±1°C;
2.8. Temperature Uniformity: ±3°C (at 400°C platform, tested with an empty furnace for 1 hour);
2.9. Detection Thermocouple: K-type thermocouple;
2.10. Exhaust Chimney: 1, DN50, with manual adjustment of exhaust port size;
2.11. Temperature Controller: FP93 temperature controller;
2.12. Number of Temperature Control Segments: 4 × 10 segments;
2.13. Number of Temperature Control Points: 1 point;
2.14. Number of Detection Points: 1 point;
2.15. Alarm Protection: Audible and visual alarms for over-temperature, open circuit, low pressure, and low water pressure;
2.16. Over-temperature Protection: Independent temperature-sensing over-temperature protector;
2.17. Motor Protection: Over-current protection for the motor;
2.18. Surface Temperature Rise: ≤35°C;
2.19. Heating Rate: ≤5°C/min;
2.20. Heating Power: 12 kW;
2.21. Insulation Power: ≤6 kW;
2.22. Reference External Dimensions: 1110mm × 1050mm × 1820mm (W × D × H, subject to actual product);
2.23. Equipment Color: Light grey.
3. Delivery List
|
Item |
Note |
Qty. |
Basic Components |
Furnace |
|
1 unit |
inspection certificate |
Certificate of major outsourced components |
1 SET |
Technical Documents |
Instructions, technical documents of the major outsourced parts, etc |
1 SET |
Key Components |
Heating Element |
|
1 SET |
Temperature Controller |
|
1 SET |
Circulation Fan |
|
1 SET |
Spare Parts |
SSR |
|
1 pcs |
Heating wire |
|
1 PC |
4. Normal Operating Conditions
1.Environmental Conditions: Temperature 0 to 40°C, humidity ≤80% RH, no corrosive gases, no strong air disturbance;
2. Ventilation System: Non-contact connection to the user's extraction system, with an extraction capacity greater than 5 m³/h;
3. Floor Requirements: Level, no significant vibration, load-bearing capacity >200 Kg/m²;
4. Power Requirements: Capacity greater than 16 kVA, 3-phase 5-wire, voltage 220/380V, frequency 50Hz. Live wires: Yellow, Green, Red; Neutral wire: Blue; Earth wire: Green/Yellow;
5. Installation Site: 2000mm × 1800mm × 3000mm (W × D × H), installation area should be greater than 4m².