Overview
The HBF105-14 Type Sintering Furnace uses silicon molybdenum rods, which are resistant to acids and alkalis, with good uniformity and simple operation. It is suitable for various laboratories and small-batch production processes.
Main Parameters
Product Model: HBF105-14 Type Sintering Furnace
External Dimensions: 850 * 900 * 1200mm (W×H×D)
Effective Chamber Size: 400 * 330 * 800mm (W×H×D)
Rated Temperature: 1300ºC
Maximum Temperature: 1500ºC
Loading Device: Corundum Mullite Loading Plate
Heating Element: Silicon Molybdenum Rod
Temperature Control Stability: ±1
Chamber Temperature Uniformity: ±5ºC
Number of Temperature Control Segments: 40 segments
Alarm Protection: Interlock between heating and furnace door operation, over-temperature, thermocouple break, and low-pressure audible and visual alarms
Surface Temperature Rise: <35ºC
Number of Temperature Control Points: 1 point
Maximum Heating Power: 23kW
Cooling Method: Natural Cooling with the Furnace
Heating Rate: ≤10ºC/min
Equipment Weight: 500Kg
Installation Dimensions: 800 * 900 * 1200mm (D×W×H)
Power Supply: Capacity greater than 30kVA, three-phase five-wire 220/380V, 50Hz
Item |
pecifications |
Quantity |
Notes |
Kiln Main Unit |
Primary heating system |
1 set |
Core equipment for thermal process
|
Setter Plate |
Material: Corundum-Mullite |
1 set |
High-temperature resistant carrier |
Installation and Usage Conditions
- The equipment must be installed indoors, with an ambient temperature of -10ºC to 45ºC. The indoor environment should be free of flammable gases, corrosive gases, and steam.
- The equipment requires a floor area of no less than 4 square meters, with a working space length of no less than 2 meters, width of no less than 2 meters, and height of no less than 3 meters. The foundation should be sturdy and level, with the furnace weighing approximately 400kg.
- The power supply for this furnace must be a three-phase five-wire AC power line, 380±10%V, 50±2%Hz. This power line should be dedicated, and large power conversion equipment such as welding machines, electric furnaces, and metal cutting machines should not be connected to this distribution.
- The equipment must be reliably grounded.